Recent alarms page alarm history page – Welch Allyn Connectivity Server - User Manual User Manual
Page 91

Directions for Use
Specifications
87
Recent Alarms page
Alarm History page
Table 13. WACS AcuityLink option: Clinician Notifier Recent Alarms page
Feature
Specification
Recent alarms shown
Active and inactive alarms, including arrhythmia events that
have been deleted from Acuity System review windows and
printouts
Situations when only a patient’s first alarm is shown
•
When a lower priority alarm occurs while another alarm
is asserting or suspended
•
When a higher priority alarm occurs while another alarm
is suspended, but the new alarm is neither a lethal
arrhythmia nor apnea alarm.
Number of alarms per patient stored in the page
Default: 25 alarms
Configurable 25-200 alarms
Recent alarms retention period
Default 4 hours
Configurable: 5 minutes-12 hours
Table 14. WACS AcuityLink option: Clinician Notifier Alarm History page
Feature
Specification
Alarms shown in Alarm History page
All of a patient’s most recent Acuity System alarms, including
arrhythmia events that have been deleted from Acuity System
review windows and printouts.
These types of alarms can be listed in the Alarm History page
without ever having appeared on the mobile device as an
alarm:
•
Alarm types not configured in the WACS Clinician Notifier
Configure Notification page to be delivered to mobile
devices
•
Low-level arrhythmia alarms (as set at Acuity System) are
not delivered to mobile devices running Clinician Notifier.
•
Short-duration alarms (shorter than the alarm delivery
holdoff variable in WACS Configure Notification page) are
not delivered to mobile devices running Clinician Notifier
•
Higher priority alarms that occur when another alarm is
suspended, but the new alarm does not interrupt the
suspend period because it is neither a lethal arrhythmia
nor an apnea alarm.
•
Lower priority alarms that occur while another alarm is
asserting or suspended
•
Multiple listings for the same alarm if the alarm’s popup
is replaced by another alarm that later ceases. The
original alarm is still active after the popup ceases.
Number of alarms per patient stored in the page
Default: 25 alarms
Configurable 25-200 alarms
Length of time alarm histories are shown after
disconnection from Acuity System
Default 96 hours
Configurable: 1-96 hours, but no longer than the Acuity
System Full Disclosure license period