Welch Allyn Connectivity Server - User Manual User Manual
Page 42

38
Configure the WACS HL7 Interface option
Welch Allyn Connectivity Server
Allow observations for multiple patients in one HL7 message
Multiple patient observations setting off
If this setting is not checked, WACS sends each patient's vital signs in a separate HL7
message.
This example shows two independent messages, each with its own acknowledgement.
Messages resulting from multiple patient observations setting off
HL7# HL7Log Initiator sending message:
MSH|^~\&|WAP^WAP||||20090129092200.893-0800||ORU^R01|200901290922008932|P|2.4
PID|1|9582173|9582173||Van Goe^Edgar^A|||||||||||||9582173
PV1|1||TEST^102^GA001150^^^^^^PropaqCS 246
OBR|1|||VITALS^Vital Signs^WAP|||20090129092000.000-0800
OBX|1|ST|Heart Rate^Heart Rate^WAP||80|^BPM||||||||20090129092000.000-0800
OBX|2|ST|Temperature^Temperature^WAP|1|98.6|^°F||||||||20090129092000.000-0800
OBX|3|ST|Temperature^Temperature^WAP|2|97.5|^°F||||||||20090129092000.000-0800
OBX|4|ST|SPO2^SPO2^WAP||97|^%||||||||20090129092000.000-0800
OBX|5|ST|CO2 (In)^CO2 (In)^WAP||0.0|^%||||||||20090129092000.000-0800
OBX|6|ST|CO2 (Ex)^CO2 (Ex)^WAP||4.9|^%||||||||20090129092000.000-0800
OBX|7|ST|RR/BR^RR/BR^WAP||12|^Br/M||||||||20090129092000.000-0800
OBX|8|ST|PVC^PVC^WAP||0|^PVC/Min||||||||20090129092000.000-0800
HL7# HL7Log Initiator received message:
MSH|^~\&|||WAP^WAP||20090129092200.893-0800||ACK^R01|200901290922008932|P|2.4
MSA|AA|200901290922008932
HL7# HL7Log Initiator sending message:
MSH|^~\&|WAP^WAP||||20090129092200.945-0800||ORU^R01|200901290922009453|P|2.4
PID|1|1567252|1567252||Bizmark^Michelle^D|||||||||||||1567252
PV1|1||TEST^103^GA005267^^^^^^PropaqCS 242
OBR|1|||VITALS^Vital Signs^WAP|||20090129092000.000-0800
OBX|1|ST|Heart Rate^Heart Rate^WAP||80|^BPM||||||||20090129092000.000-0800
OBX|2|ST|Temperature^Temperature^WAP|1|98.6|^°F||||||||20090129092000.000-0800
OBX|3|ST|Temperature^Temperature^WAP|2|97.5|^°F||||||||20090129092000.000-0800
OBX|4|ST|SPO2^SPO2^WAP||97|^%||||||||20090129092000.000-0800
OBX|5|ST|RR/BR^RR/BR^WAP||12|^Br/M||||||||20090129092000.000-0800
OBX|6|ST|PVC^PVC^WAP||0|^PVC/Min||||||||20090129092000.000-0800
HL7# HL7Log Initiator received message:
MSH|^~\&|||WAP^WAP||20090129092200.945-0800||ACK^R01|200901290922009453|P|2.4
MSA|AA|200901290922009453