Metrohm 757 VA Computrace User Manual

Page 89

background image

5.2 Working method specifications

757 VA Computrace – Software

79

rameters are not displayed if

Batch with solution

exchange

is selected for

Technique

.

No. [ 1...5 ; 0 ]

Number of addition solution to be used for man-
ual or automatic addition. For automatic addi-
tions, this number is also the Dosimat number. If
a mixed standard solution is used, the number of
this solution must be entered for each substance
present in the mixed solution.

Conc. [ > 0 ; 0 ]

Value for concentration of addition solution.

Unit [ µg/L...g/µL ; g/L ]

Unit for concentration of addition solution.

Volume (mL) [ > 0.01 mL / var ; 0 mL ]

Addition volume. For entering variable addition
volumes, click the button to open the

EDIT

VARIED ADDITION

window for the selected sub-

stance (details see Variable addition). In this
case,

var

is entered into the field instead of a

fixed value. This field only appears once for solu-
tions with the same number (mixed standards)
and it is not displayed if

Batch with solution ex-

change

is selected for

Technique

or if

0

is entered

as solution number.

Click this button to open the

CELL

CONCENTRATIONS

window for entering the con-

centrations of the solutions used for standard
addition or the recording of calibration curves for
the selected substance (details see Concentra-
tions of calibration solutions
). This button is only
displayed if

Batch with solution exchange

is se-

lected for

Technique

.

Regression technique [ see below ; Linear Regression ]

Selection of regression technique:

Linear

Regression

The regression is calculated with a straight line.

Nonlinear

Regression

The regression is calculated with a nonlinear cur-
ve. This option is only available if

Record calibrati-

on curve

is selected for

Calibration

.

Linear Regression (through Zero)

The regression is calculated with a straight line
forced to the zero point. This option is only avai-
lable if

Record calibration curve

is selected for

Ca-

libration

.