Stand-by potential – Metrohm 757 VA Computrace User Manual
Page 58
3 General settings for exploratory and determination mode
757 VA Computrace – Software
48
• During the
equilibration time
, the start potential of the
sweep is applied to the electrode.
Cleaning potential (V)
[ -5...+5 V ; -0.1 V ]
Voltage applied to the electrodes during the
Cleaning time
.
Cleaning time (s)
[ 0...80600 s ; 0 s ]
Time during which the
Cleaning
potential
is applied to the
electrodes.
Deposition potential (V)
[ -5...+5 V ; -0.9 V ]
Voltage applied to the electrodes during the
Deposition time
.
Deposition time (s)
[ 0...80600 s ; 60 s ]
Time during which the
Deposition
potential
is applied to the
electrodes.
Equilibration time (s) [ 0...80600 s ; 5 s ]
Waiting time before starting the sweep with stirrer switched
off and start potential applied to the electrodes.
Stand-by potential
The
stand-by potential can be applied to the electrodes at the
end of the measurement. It remains in force until it is switched off
manually in the
COMPUTRACE CONTROL
window or until a new
voltage is applied to the electrodes in the next measurement.
Cell off after measurement
[ on, off ; on ]
Enable/disable the switching off of the voltage applied to the
electrodes after measurement.
Stand-by potential (V)
[ -5...+5 V ; -0.1 V ]
Voltage to be applied to the electrodes after measurement if
the
Cell off after measurement
box is set to
off
.