Perform a linearity test with the dummy cell – Metrohm 757 VA Computrace User Manual
Page 167
![background image](/manuals/586553/167/background.png)
7.9 Diagnostic procedures
757 VA Computrace – Software
157
8. Select
SMDE
and click on
.
9. A small drop is formed and removed in intervals. The drop
size has to be reproducible and the drop has to be removed
in every interval.
10. If the drop is not removed reproducibly, adjust the tapper
mechanism (see Hardware Manual).
11. Select
HMDE
and click on
.
12. A small new drop is formed and stays at the end of the capil-
lary. The drop should remain for at least 1 min. Otherwise the
capillary and/or the needle should be changed (see Hardware
Manual).
Perform a linearity test with the dummy cell
For testing the linearity of current measurement, the dummy cell
of the 757 VA Computrace Stand is used with the test method
Test757_L.mth
. Proceed as follows:
1. Click on
or
MAIN WINDOW
/ Mode / Determination
.
2. Click on
or
MAIN WINDOW
/ Window / Working method
specification
to open the
WORKING METHOD SPECIFICATIONS
window.
3. Click on
or
MAIN WINDOW
/ File / Load method
.
4. Select the method file
Test757_L.mth
in the
OPEN
window and
click
. The method is loaded into the
WORKING METHOD
SPECIFICATIONS
window.
5. Connect the dummy cell at the 757 VA Computrace Stand: At-
tach electrode cable AE to clamping screw
AE, attach elec-
trode cable RE to clamping screw
RE, attach electrode cable
WE to clamping screw
WE-L.
6. Click on
or
MAIN WINDOW
/ Window / Monitor
to open the
MONITOR
window.
7. Start the measurement by clicking the
icon in the
MAIN
WINDOW
or the
button in the
MONITORING
window.
8. Enter the
Sample ID
(used as part of the determination file
name) in the
PLACE SAMPLE
window and click
.
9. At the end of the measurement, a curve is printed out. This
curve should satisfy the following conditions:
• The plotted diagonal must be straight.
• At -200 mV, the current should be -1.6…-2.4 µA.
• At +200 mV, the current should be +1.6…+2.4 µA.