7.6 Standard addition technique
757 VA Computrace – Software
147
15. Enter the
Sample ID
(used as part of the determination file
name) in the
PLACE SAMPLE
window and click
.
16. Each time a standard addition is required in the
MANUAL AD-
DITION
window, add the standard addition solution using a
pipette.
Use manual standard addition with solution exchange
In the manual standard addition with solution exchange, a new
sample solution is used for every standard addition. Proceed as
follows:
1. Click on
or
MAIN WINDOW
/ Mode / Determination
.
2. Click on
or
MAIN WINDOW
/ Window / Working method
specification
to open the
WORKING METHOD SPECIFICATIONS
window.
3. Load the desired method into the
WORKING METHOD SPECIFI-
CATIONS
window (see How to load a method).
4. Select
Standard addition
in the
Calibration
field of the
WORKING
METHOD SPECIFICATIONS
window.
5. Select
Manual
in the
Addition
field of the
WORKING METHOD
SPECIFICATIONS
window.
6. Select
Batch with solution exchange
in the
Technique
field of the
WORKING METHOD SPECIFICATIONS
window.
7. Click
to open the
EDIT WORKING METHOD PA-
RAMETERS
window.
8. Select the
Determination
tab (see section 5.2) and enter
Sam-
ple identifier
,
Sample amount
,
Cell volume
, and the number of
spiked solutions in the
No. of cells
field.
9. Select the
Substances
tab (see section 5.2) and make sure
that for each substance entered in the table the concentra-
tions of the spiked sample solutions are defined in the
CELL
CONCENTRATIONS
window which is opened by clicking on the
button.
10. Close the
EDIT WORKING METHOD PARAMETERS
window by
clicking
.
11. Place the sample solution in the measuring vessel at the 757
VA Computrace Stand.
12. Click on
or
MAIN WINDOW
/ Window / Monitor
to open the
MONITOR
window.