beautypg.com

Performing exploratory measurements – Metrohm 757 VA Computrace User Manual

Page 69

background image

4.2 Exploratory specification window

757 VA Computrace – Software

59

Performing exploratory measurements

Measurements in the exploratory mode can be performed using
the following icons (in the

MAIN WINDOW

) or buttons (in the

EX-

PLORATORY SPECIFICATION

window):

Start measurement
The operation sequence (see section 3.4) de-
fined in the

EXPLORATORY SPECIFICATION

win-

dow is started. Each step of the operation se-
quence is listed in the first line of the status win-
dow beside the

button.

For the running voltage sweep there is a live dis-
play in the

EXPLORATORY CURVES

window with

automatic scaling of the axes. Manual rescaling
can be done by pressing the button.

The red light at the left side of the

button

indicates a current overload. In this case, stop
the measurement and change the measurement
parameters.

In the second line of the status window com-
ments or error messages concerning the running
measurement are displayed.

A running measurement can be stopped, inter-
rupted and continued. Each step in the operation
sequence can be abbreviated by clicking the

button.

Stop measurement
Stop running measurement immediately.

Hold measurement
Interrupt running measurement.

Continue measurement
Continue an interrupted measurement.

Next step