7 How to ...?
757 VA Computrace – Software
152
13. Start the measurement by clicking the
icon in the
MAIN
WINDOW
or the
button in the
MONITORING
window.
14. Enter the
Calibration curve id
(used as part of the determina-
tion file name) in the
START CALIBRATION
window and click
.
15. Each time a solution exchange is required in the
BATCH SO-
LUTION EXCHANGE
window, replace the solution measured by
the next calibration solution.
Record calibration curve automatically
This method is used for preparing different calibration solutions
by adding automatically several times a concentrated standard
solution to the measuring solution using 665 or 765 Dosimats.
Proceed as follows:
1. Install Dosimats to the 757 VA Computrace Stand (see How to
install Dosimats).
2. Click on
or
MAIN WINDOW
/ Mode / Determination
.
3. Click on
or
MAIN WINDOW
/ Window / Working method
specification
to open the
WORKING METHOD SPECIFICATIONS
window.
4. Load the desired method into the
WORKING METHOD SPECIFI-
CATIONS
window (see How to load a method).
5. Select
Record calibration curve
in the
Calibration
field of the
WORKING METHOD SPECIFICATIONS
window.
6. Select
Automatic
in the
Addition
field of the
WORKING METHOD
SPECIFICATIONS
window.
7. Click
to open the
EDIT WORKING METHOD PA-
RAMETERS
window.
8. Select the
Determination
tab (see section 5.2) and enter the
Cell volume
and the number of additions in the
No. of additions
field.
9. Select the
Substances
tab (see section 5.2) and make sure
that for each substance entered in the table the number of the
single or mixed standard solution, its concentration and its
volume is defined. The
No.
of the standard solution must be
identical to the number of the Dosimat used for automatic
addition of this solution.
10. If the addition of the standard solution should be done with
variable addition volumes, click the button in the
Volume
column to open the
EDIT VARIED ADDITION
window, enter the