7.7 Calibration curve technique
757 VA Computrace – Software
151
15. Enter the
Calibration curve id
(used as part of the determina-
tion file name) in the
START CALIBRATION
window and click
.
16. Each time an addition is required in the
MANUAL ADDITION
window, add the standard solution using a pipette.
Record calibration curve manually with solution exchange
This method is used for recording a calibration curve using differ-
ent calibration solutions of known concentration. Proceed as fol-
lows:
1. Click on
or
MAIN WINDOW
/ Mode / Determination
.
2. Click on
or
MAIN WINDOW
/ Window / Working method
specification
to open the
WORKING METHOD SPECIFICATIONS
window.
3. Load the desired method into the
WORKING METHOD SPECIFI-
CATIONS
window (see How to load a method).
4. Select
Record calibration curve
in the
Calibration
field of the
WORKING METHOD SPECIFICATIONS
window.
5. Select
Manual
in the
Addition
field of the
WORKING METHOD
SPECIFICATIONS
window.
6. Select
Batch with solution exchange
in the
Technique
field of the
WORKING METHOD SPECIFICATIONS
window.
7. Click
to open the
EDIT WORKING METHOD PA-
RAMETERS
window.
8. Select the
Determination
tab (see section 5.2) and enter the
Cell volume
and the number of calibration solutions in the
No.
of cells
field.
9. Select the
Substances
tab (see section 5.2) and make sure
that for each substance entered in the table the concentra-
tions of the calibration solutions are defined in the
CELL
CONCENTRATIONS
window which is opened by clicking on the
button.
10. Close the
EDIT WORKING METHOD PARAMETERS
window by
clicking
.
11. Place the first calibration solution in the measuring vessel at
the 757 VA Computrace Stand.
12. Click on
or
MAIN WINDOW
/ Window / Monitor
to open the
MONITOR
window.