Metrohm 757 VA Computrace User Manual
Page 159
7.6 Standard addition technique
757 VA Computrace – Software
149
13. If the standard addition should be done with variable addition
volumes, click the button in the
Volume
column to open
the
EDIT VARIED ADDITION
window, enter the variable addition
volumes in the
Addition
fields, and close this window by click-
ing
.
14. Close the
EDIT WORKING METHOD PARAMETERS
window by
clicking
.
15. Place the sample solution in the measuring vessel at the 757
VA Computrace Stand.
16. Click on
or
MAIN WINDOW
/ Window / Monitor
to open the
MONITOR
window.
17. Start the measurement by clicking the
icon in the
MAIN
WINDOW
or the
button in the
MONITORING
window.
18. Enter the
Sample ID
(used as part of the determination file
name) in the
PLACE SAMPLE
window and click
.