7.9 Diagnostic procedures
757 VA Computrace – Software
159
Perform a GLP test
The requirements of GLP (Good Laboratory Practice) include the
periodic testing of analytical instruments with regard to repro-
ducibility and accuracy using
standard operating procedures.
Metrohm suggests the procedure described below as the stan-
dard operating procedure for testing the 757 VA Computrace
Stand:
1. Click on
or
MAIN WINDOW
/ Mode / Determination
.
2. Click on
or
MAIN WINDOW
/ Window / Working method
specification
to open the
WORKING METHOD SPECIFICATIONS
window.
3. Perform the linearity test with the dummy cell (see How to per-
form a linearity test).
4. Perform the peak test with the dummy cell (see How to per-
form a peak test).
5. Click on
or
MAIN WINDOW
/ File / Load method
.
6. Select the method file
Test Pb in standard solution.mth
in the
OPEN
window and click
. The method is loaded into the
WORKING METHOD SPECIFICATIONS
window.
7. Add 20 mL ultrapure water to the empty measuring vessel at
the 757 VA Computrace Stand.
8. Add 0.5 mL potassium chloride c(KCl) = 3 mol/L (Metrohm
No. 6.2308.020) to the measuring vessel.
9. Click on
or
MAIN WINDOW
/ Window / Monitor
to open the
MONITOR
window.
10. Start the measurement by clicking the
icon in the
MAIN
WINDOW
or the
button in the
MONITORING
window.
The
PLACE SAMPLE
window appears.
11. Add 100 µL Pb ion standard solution
ρ
(Pb) = 1 g/L (Metrohm
No. 6.2301.100) into the measuring vessel and click the
button.
12. The sample solution is measured three times. Then the
MAN-
UAL ADDITION
window appears.
13. Use a pipette to add 100 µL Pb ion standard solution
ρ
(Pb) =
1 g/L (Metrohm No. 6.2301.100) into the measuring vessel
and click the
button.
14. The sample solution spiked with standard addition solution is
measured three times. Then the
MANUAL ADDITION
window
appears.