Evaluate signal peaks automatically, Evaluate signal peaks manually – Metrohm 757 VA Computrace User Manual
Page 146
7 How to ...?
757 VA Computrace – Software
136
7. Set pretreatment parameters for electrodes (see Pretreatment,
section 3.4).
8. Set
Sweep
parameters of the selected VA measurement mode
(see VA measurement modes, section 3.2).
9. If desired, set
Stand-by potential
to be applied after measure-
ment (see Stand-by potential, section 3.4).
10. If the recorded signal file should be automatically saved at the
end of the measurement, click on
MAIN WINDOW / Settings /
General settings
and enable the
Auto save determination and sig-
nal
option in the
General
tab.
11. Start the measurement by clicking the
icon or the
button (see Performing measurements, section 4.2).
Evaluate signal peaks automatically
1. Select the desired signal file in the
Signal
field of the
EX-
PLORATORY SPECIFICATION
window. The selected signal
curve is shown with the
Selected signal properties
.
2. Click on
EXPLORATORY SPECIFICATION
/ Signal / Peak search
.
The
PEAK SEARCH
window is opened.
3. Set the parameters
Reverse peak
,
Reverse sweep
(for CV only),
Minimum peak width
,
Smooth factor
,
Minimum peak height
, and
Scope
for peak evaluation (see Peak search, section 4.2).
4. Click the
button. The calculated baselines and
peak maximum positions are displayed in the
EXPLORATORY
CURVES
window. The evaluation results are displayed in the
table of results in the
EXPLORATORY SPECIFICATION
window.
5. If no peaks are found, try to modify the peak search parame-
ters
Minimum peak width
,
Smooth factor
, and
Minimum peak
height
for the automatic signal peak evaluation or switch to
the manual signal peak evaluation.
Evaluate signal peaks manually
1. Select the desired signal file in the
Signal
field of the
EX-
PLORATORY SPECIFICATION
window. The selected signal
curve is shown with the
Selected signal properties
.
2. Click on
EXPLORATORY SPECIFICATION
/ Signal / Peak search
.
The
PEAK SEARCH
window is opened.
3. Check the
Manual
option.
4. Set the parameters
Reverse peak
,
Reverse sweep
(for CV only),
Minimum peak width
,
Smooth factor
, and
Minimum peak height
for peak evaluation (see Peak search, section 4.2).