7 How to ...?

757 VA Computrace – Software

136

7. Set pretreatment parameters for electrodes (see Pretreatment,

section 3.4).

8. Set

Sweep

parameters of the selected VA measurement mode

(see VA measurement modes, section 3.2).

9. If desired, set

Stand-by potential

to be applied after measure-

ment (see Stand-by potential, section 3.4).

10. If the recorded signal file should be automatically saved at the

end of the measurement, click on

MAIN WINDOW / Settings /

General settings

and enable the

Auto save determination and sig-

nal

option in the

General

tab.

11. Start the measurement by clicking the

icon or the

button (see Performing measurements, section 4.2).

Evaluate signal peaks automatically

1. Select the desired signal file in the

Signal

field of the

EX-

PLORATORY SPECIFICATION

window. The selected signal

curve is shown with the

Selected signal properties

.

2. Click on

EXPLORATORY SPECIFICATION

/ Signal / Peak search

.

The

PEAK SEARCH

window is opened.

3. Set the parameters

Reverse peak

,

Reverse sweep

(for CV only),

Minimum peak width

,

Smooth factor

,

Minimum peak height

, and

Scope

for peak evaluation (see Peak search, section 4.2).

4. Click the

button. The calculated baselines and

peak maximum positions are displayed in the

EXPLORATORY

CURVES

window. The evaluation results are displayed in the

table of results in the

EXPLORATORY SPECIFICATION

window.

5. If no peaks are found, try to modify the peak search parame-

ters

Minimum peak width

,

Smooth factor

, and

Minimum peak

height

for the automatic signal peak evaluation or switch to

the manual signal peak evaluation.

Evaluate signal peaks manually

1. Select the desired signal file in the

Signal

field of the

EX-

PLORATORY SPECIFICATION

window. The selected signal

curve is shown with the

Selected signal properties

.

2. Click on

EXPLORATORY SPECIFICATION

/ Signal / Peak search

.

The

PEAK SEARCH

window is opened.

3. Check the

Manual

option.

4. Set the parameters

Reverse peak

,

Reverse sweep

(for CV only),

Minimum peak width

,

Smooth factor

, and

Minimum peak height

for peak evaluation (see Peak search, section 4.2).