beautypg.com

6 sample table – Metrohm 757 VA Computrace User Manual

Page 118

background image

5 Determination mode

757 VA Computrace – Software

108

5.6 Sample

table

The

SAMPLE TABLE

window shows the sample data of the loaded

sample table.

Pos. [ 1, 3, 5 ... 127; read only ]

Position of the sample on the sample rack of the
813 Compact Autosampler. Only odd numbers
are displayed, since for every sample a vessel
containing rinsing solution must be placed at the
even sample rack positions.

Sample ID [ 16 characters ; "sample" ]

Identification for sample.

Amount [ > 0 ; 10 ]

Amount of sample added to the measuring ves-
sel.

Unit [ mL, g ; mL ]

Selection of unit for sample amount.

Cell volume (mL) [ > 0 mL ; 10 mL ]

Total volume of solution (sample + auxiliary solu-
tion, e.g. buffer) in the measuring vessel at the
start of the determination. The sample concentra-
tions Mass conc. calculated refer to this
cell volume.