beautypg.com

Hardware settings for dosimats – Metrohm 757 VA Computrace User Manual

Page 17

background image

1.3 Installation

757 VA Computrace – Software

7

5. Make hardware settings for the 813 Compact Autosampler

(see Automation, section 2.7).

6. Make hardware settings for Dosimats (see Hardware settings

for Dosimats).

7. Define the addition or predose solution in the

DOSIMATS

win-

dow (see Dosimats, section 5.2).


Hardware settings for Dosimats

1. Switch on PC and start operating system 95.
2. Start the VA Computrace software by double-clicking the pro-

gram icon or starting the

ct757.exe

file.

3. Login without entering

Name

and

Password

. Click on

Recalcu-

late only

.

4. Select

MAIN WINDOW / Settings / General Settings.

Open the

Hardware

tab in the

GENERAL SETTINGS

window:

5. For each 665 or 765 Dosimat connected to the remote inter-

face of the 757 VA Computrace Stand, check the

Dosimat no.

checkbox and enter the

Volume

of the exchange unit installed

on the Dosimat and the

Dose rate

of the Dosimat.