5 Determination mode

757 VA Computrace – Software

76

Cell volume (mL) [ > 0 mL ; 10 mL ]

Total volume of solution (sample + auxiliary solu-
tion, e.g. buffer) in the measuring vessel at the
start of the determination. The sample concentra-
tions Mass conc. calculated refer to this cell
volume.

Measure blank [ on, off ; off ]

Measure a blank solution before sample deter-
mination. The blank curve is then automatically
subtracted from all subsequent measured
curves. This background compensation is mainly
used to reduce interference due to the support-
ing electrolyte. Such interference includes both
the presence of the analyte (blank value) and that
of foreign substances electroactive in the same
range. The blank curve is only recorded once (no
replications).

No. of blanks [ 1...5 ; 1 ]

Number of measurements to determine the blank
curve. If the blank solution is measured several
times, a mean blank curve is determined from
the different measurements.

Blank purge time [ 0...80600 s ; 300 s ]

Time of inert gas purging before measurement of
the blank solution.

Addition purge time [ 0...80600 s ; 10 s ]

Time of inert gas purging for all measurements
after determination of the sample (standard addi-
tion) or of the first calibration solution if

Batch

is

selected for

Technique

(for the first measurement,

the

Initial purge time

is used).

Cell purge time [ 0...80600 s ; 10 s ]

Time of inert gas purging after solution exchange
if

Batch with solution exchange

is selected for

Technique

(for the first measurement, the

Initial

purge time

is used).

No. of additions [ 0...28 ; 2 ]

Number of additions of standard addition solu-
tions or calibration solutions if

Batch

is selected

for

Technique

.

No. of cells [ 0...28 ; 2 ]

Number of solutions to be measured if

Batch with

solution exchange

is selected for

Technique

.

Calibration curve [ path + file name ; ]

Selection of the determination file with the re-
corded calibration curves to be used if

Sample

with calibration curve

is selected for

Calibration

.