2.15 Storing determination results and silo calculations

784 KFP Titrino

41

2.15 Storing determination results and silo calcula-
tions

2.15.1 Storing determination results

If the sample-specific data of the silo memory should be kept after the determination and

supplemented by results, the following entries are necessary:

1. In the method under

Assignment of the determination results to C24 and/or C25:

2. In the silo memory, (when the silo memory is switched in):

"save lines: on"

Assignment of determination results

The determination results are assigned in key .

The display texts of the Titrino are shown on the left

side.

>silo calculations

C24=
C25=

Silo calculations

Assignment to C24 (RSX, EPX, CXX)

Calculated results (RSX), endpoints (EPX) or variables

CXX can be stored as C24.

Same procedure for C25.

Important:

Ensure that there is still sufficient space for storing the results C24 and C25. (In the report

the number of free bytes is shown.) Result name,

value and unit are stored. The memory requirements can be estimated as follows:

Result with text (8 characters) and unit (5 characters):

32 bytes

Measured value C40, value without unit:

22 bytes

DEF

(

2

def
>formula

>silo calculations

>common variables
>report
>mean