Metrohm 784 KFP Titrino User Manual
Page 44

2.15 Storing determination results and silo calculations
784 KFP Titrino
41
2.15 Storing determination results and silo calcula-
tions
2.15.1 Storing determination results
If the sample-specific data of the silo memory should be kept after the determination and
supplemented by results, the following entries are necessary:
1. In the method under
Assignment of the determination results to C24 and/or C25:
2. In the silo memory,
"save lines: on"
Assignment of determination results
The determination results are assigned in key
The display texts of the Titrino are shown on the left
side.
>silo calculations
C24=
C25=
Silo calculations
Assignment to C24 (RSX, EPX, CXX)
Calculated results (RSX), endpoints (EPX) or variables
CXX can be stored as C24.
Same procedure for C25.
Important:
Ensure that there is still sufficient space for storing the results C24 and C25. (In the report
value and unit are stored. The memory requirements can be estimated as follows:
Result with text (8 characters) and unit (5 characters):
32 bytes
Measured value C40, value without unit:
22 bytes
DEF
(
2
def
>formula
>silo calculations
>common variables
>report
>mean