2.7 Result calculations

784 KFP Titrino

26

RS1 text RS1

RS1 decimal places 2

RS1 unit: %

RS1 limit control: OFF

RS1 low lim. 0.0
RS1 up lim. 0.0

RS1 L13 output: OFF

Text for result output (up to 8 characters)

Text input see page 8.

Number of decimal places for result (0...5)

Selection of result unit (%, ppm, g/L, mg/mL, mol/L,

mmol/L, g, mg, mL, mg/pc, s, mL/min, no unit or up to

6 characters).

Limit control for the result (ON, OFF)

The limits are checked each time a result is calculated.

If "ON" has been set:

Lower limit (0.0...999 999)

Upper limit (0.0...999 999)

Sets line L13 of the remote socket (OFF, active, pulse)

if the result lies outside the limits.

Enter next formula, e.g. for RS2.

Meaning of the calculation variables CXX:

C00

Sample size, see page 37.

C01...C19

Method specific operands, see page 27. They are stored with the

method in the method memory.

C21...C23

Sample specific operands, see page 37ff.

C26, 27

Mean values from silo calculations, see page 41ff.

C30...C39

Common variables.

C40

Initial measured value of the sample.

C41

End volume.

C42

Determination time.

C43

Volume drift for KFT with conditioning.

C44

Temperature.

C45

Dispensed start volume.