4.1 Error and special messages
784 KFP Titrino
97
4 Error messages and Troubleshooting
Data transfer inoperative See measures on page 95.
4.1 Error and special messages
XXX bytes missing
For the storage of a method or a silo line XXX bytes are
missing.
Remedy:
fewer silo lines.
check electrode
With polarized electrodes. There is a break or short circuit.
Possible causes and rectification of the fault:
- the electrode is not plugged in
⇒
plug it in
- the electrode is not immersed in the solution
⇒
immerse it
- the electrode is defective
⇒
use new electrode.
- the electrode cable is defective
⇒
use new cable.
The electrode test can be switched off under the
key.
Exit: Rectify fault or
check exchange unit!
The Exchange Unit is not mounted (properly).
Exit: Mount Exchange Unit (properly) so that the coupling
engages or
check remote box
The Remote Box is not (correctly) connected or the Remote
Box is connected but not activated under the
Exit: Connect Remote Box (correctly) and set "Remote Box:
ON" under
off/on.
division by zero
The result could not be calculated as a divisor in the formula
was equal to zero.
Exit: Enter appropriate value.
error 36
Parity
Exit:
same on both instruments
error 37
Stop bit
Exit:
same on both instruments
error 38
Overrun error. At least 1 character could not be read.
Exit:
error 39
Overflow of the receive buffer of the GP Titrino (> 128
characters).
Exit:
error 42
CTS=OFF
Handshake unsatisfactory for more than 1 s.
Exit:
to receive?
error 43
The transmission of the GP Titrino has been
interrupted with XOFF for at least 6 s.
Exit: