Metrohm 784 KFP Titrino User Manual
Page 22

2.6 Parameters, key
784 KFP Titrino
19
req.smpl size: OFF
cond.
limit smpl size: OFF
cond.
low lim. 0.0
cond.
up lim. 999999
cond.
oven: no
cond.
activate pulse: OFF
cond.
Request of sample size after start of titration (value,
unit, all, OFF)
"all" the value and the unit will be requested.
Limiting value check for sample size (ON, OFF)
With "ON" the error message "sample size out." ap-
pears if the entry is outside the set limits. The limiting
values are shown in the display window.
The absolute value of the limit is checked during sam-
ple size input and during the calculation of the results.
If "ON" has been set:
Lower limit for sample size (0.0...999 999)
Upper limit for sample size (0.0...999 999)
Connected oven (COM1, COM2, no)
COM of the Titrino to which the oven is connected.
If an oven is connected via RS232 an inquiry will be
made for the oven results and these will be inserted
into the result report of the Titrino. The report output on
the oven must be switched OFF (see also page 115).
Set "no" if no oven has been connected or if you have
not connected the oven to Titrino the via RS232 inter-
face.
Pulse output on I/O line L6 (L6, pin 1) of the remote
socket (first, all, cond., OFF) e.g. start of a Dosimat,
see page 126.