beautypg.com

3 user methods – Metrohm 784 KFP Titrino User Manual

Page 135

background image

6.3 User methods

784 KFP Titrino

132

6.3.4 Blank determination for KFT "Blank_KF"

This method can be generally used for KF

blank determinations. For work with a KF

oven, you need to enter an extraction time in

key , ">titration parameters".

The mean value out of 3 determinations is as-

signed to the common variable C38.

Electrode:

Double Pt-electrode 6.0338.100 at measuring

input "Pol".

Titrating agent:

One-component or two-component Karl-

Fischer reagent, ready to use.

Solvent:

20 mL Methanol or a specific solvent, condi-

tioned.

Sample:

E.g. 1.000 mL methanol (as used for extrac-

tion) or another solvent.

References:

G. Wieland, Water determination by Karl

Fischer Titration, GIT Verlag, Darmstadt, Ger-

many

HYDRANAL® Practical Course, Water rea-

gents for Karl-Fischer-Titration according to

Eugen Scholz, Riedel de Haën, Seelze, Ger-

many

METROHM Application Bulletin No. 77:

Karl Fischer Water Determinations

C38 is the common variable for the KF blank value.

'pa

784 KFP Titrino 784.0010

date 1999-08-18 time 18:01 0

KFT Ipol Blank_KF

parameters

>control parameters

EP at U 250 mV

dynamics 100 mV

max.rate max. ml/min

min.volume incr. min.

µl

stop crit: drift

stop drift 20

µl/min

>titration parameters

titr.direction: -

pause 1 0 s

start V: OFF

pause 2 0 s

extr.time 0 s

I(pol) 50

µA

electrode test: OFF

temperature 25.0 °C

time interval 2 s

>stop conditions

stop V: abs.

stop V 99.99 ml

filling rate max. ml/min

>statistics

status: ON

mean n= 3

res.tab: original

>preselections

conditioning: ON

display drift: ON

drift corr: OFF

req.ident: OFF

req.smpl size: OFF

limit smpl size: OFF

oven: no

activate pulse: OFF

------------

'fm

784 KFP Titrino 784.0010

date 1999-08-18 time 18:01 0

KFT Ipol Blank_KF

>calculations

Blank=EP1;4;ml

------------

'de

784 KFP Titrino 784.0010

date 1999-08-18 time 18:02

KFT Ipol Blank_KF

def

>formula

Blank=EP1

RS1 text Blank

RS1 decimal places 4

RS1 unit: ml

RS1 limit control: OFF

>silo calculations

match id: OFF

>common variables

C38=MN1

>report

report COM1:full;

>mean

MN1=RS1

>temporary variables

------------