beautypg.com

6 measuring input test: polarizer test – Metrohm 784 KFP Titrino User Manual

Page 107

background image

4.2 Diagnosis

784 KFP Titrino

104

6 Measuring input test: Polarizer test

With the aid of the "767.0010 calibrated reference for mV, pH,

Ω, µS, °C" you can check the measuring input

"Pol".

If a Remote Box is connected:

Deactivate the Remote Box (key , >peripheral units, Remote Box: off"). Switch the Titrino off and

screw off the Remote Box. Switch the Titrino on again (so that the new configuration will be recognized).

Press <7>.

polarizer test

Press .

dummy resistor 14.3 kΩ

Screw off electrode cable (6.2104.020) and insert in socket 5 of the 767 (the cover remains
closed on the 767) or connect resistor switch box (14.3 k

) using 3.496.5070 cable to ‘Pol’

socket.

Press .

Test sequence:

1. An asterisk flashes during the test.
2. In case of an error an error message appears. (If for example the switch-box is not

connected, error 100 appears).

3. If no error is found, after about 15 s display shows

polarizer test o.k.

Press .

diagnose press key 0...9

Remove 767 or cable and resistor switch-box.