beautypg.com

Metrohm 784 KFP Titrino User Manual

Page 13

background image

2.4 Configuration, key

784 KFP Titrino

10

system test report: OFF

System test report printout (ON, OFF)

With "ON" the report of the system test is printed out

after the Titrino has been switched on, see also

page 139.

>peripheral units

send to COM1: IBM

send to COM2: IBM

man.reports to COM: 1

balance: Sartorius

stirrer control: OFF

remote box: OFF

keyboard: US

Settings for peripheral units

Selection of printer (Epson, Seiko, Citizen, HP, IBM) at

the Titrino COM1

"Epson", for Epson

"Seiko", e.g. for DPU-414

"Citizen", e.g. for iDP 562 RS

"HP" e.g. for Desk Jet types. Place curves always at the

beginning of a page as you cannot have them over 2

pages.

"IBM" for all printers with IBM character set Table 437

and IBM graphics, as well as for the data transmission

to a computer or a data system.

COM of Titrino for the output of manually triggered

reports (1, 2, 1&2)

Manually triggered reports (e.g. with ....).

Exception : These reports are

outputted on the COM as defined in the method.

Selection of balance (Sartorius, Mettler, Mettler AT,

AND, Precisa)

Sartorius:

Models MP8, MC1

Mettler:

Models AM, PM and balances with 011,

012, and 016 interfaces

Mettler AT: Model AT

AND:

Models ER-60, 120, 180, 182, FR-200, 300

and FX-200, 300, 320

Precisa:

Models with RS232C interface

Automatic switching ON/OFF of the stirrer in the titration

sequence (ON, OFF)

If stirrer control is ON, the stirrer will be switched on at

the beginning and switched OFF at the end of a de-

termination. For KFT with conditioning the stirrer will be

switched off in the inactive state.

For stirrer control the red switch on the stirrer unit must

be ON.

Connection of a remote box (ON ,OFF)

To the remote socket for PC keyboard and barcode

reader, see page 118.

If "ON" has been set:

Type of PC keyboard (US, German, French, Spanish,

Swiss.)

The PC keyboard is used as an input aid, see

page 119.