4.1 Error and special messages
784 KFP Titrino
99
stop V reached
The determination has been stopped as the stop volume has
been reached.
system error 3
The instrument adjustment data have been overwritten.
Exit:
message appears each time the instrument is switched on
until it has been readjusted (Metrohm service).
system error 14
No communication between the Titrino and the connected
Remote Box.
Possible causes:
. The Remote Box was connected when the Titrino was
running
. Titrino has a fault.
. Remote Box has a fault.
Remedy: Set under
Box: OFF", switch off Titrino, take away Remote Box and
switch on Titrino. Contact Metrohm service.
time-out PC keyboard
A connected PC keyboard has been used to call up an ad-
dress (e.g.
rupted.
Possible causes:
. Remote Box has a fault.
. PC keyboard has a fault.
Exit: Correct fault and switch Titrino off/on.
transmission error
With a Remote Box connected characters are received which
cannot be interpreted.
Possible causes:
. Wrong key combination has been pressed.
. Wrong PC keyboard has been selected.
. The barcode reader supplies garbled characters.
. The Remote Box has a fault.
Exit: Rectify fault and switch Titrino off/on.
validate instrument
Validation interval has elapsed.
Exit:
wrong sample
With KFT with preset titration direction, the first measured
value is outside the end point.