4.1 Error and special messages

784 KFP Titrino

99

stop V reached

The determination has been stopped as the stop volume has

been reached.

system error 3

The instrument adjustment data have been overwritten.

Exit: . Default adjustment data are set. The error

message appears each time the instrument is switched on

until it has been readjusted (Metrohm service).

system error 14

No communication between the Titrino and the connected

Remote Box.

Possible causes:

. The Remote Box was connected when the Titrino was

running

. Titrino has a fault.

. Remote Box has a fault.

Remedy: Set under , >peripheral units, "Remote

Box: OFF", switch off Titrino, take away Remote Box and

switch on Titrino. Contact Metrohm service.

time-out PC keyboard

A connected PC keyboard has been used to call up an ad-

dress (e.g. ) and the connection has then been inter-

rupted.

Possible causes:

. Remote Box has a fault.

. PC keyboard has a fault.

Exit: Correct fault and switch Titrino off/on.

transmission error

With a Remote Box connected characters are received which

cannot be interpreted.

Possible causes:

. Wrong key combination has been pressed.

. Wrong PC keyboard has been selected.

. The barcode reader supplies garbled characters.

. The Remote Box has a fault.

Exit: Rectify fault and switch Titrino off/on.

validate instrument

Validation interval has elapsed.
Exit: or new start.

wrong sample

With KFT with preset titration direction, the first measured

value is outside the end point.