4.2 Diagnosis

784 KFP Titrino

106

7.2 Test of the remote box

The remote box needs to be connected before the Titrino is switched on. Insert 3.496.8550 test plug in the re-

mote box.

Press <2>.

reset remote box?

Press .

Test sequence:

1. In case of an error an error message is displayed (for example

error 490 01HEX).

2. If no error is found, after about 1 s display shows

remote box test o.k.

Press

. The dialogue switches automatically to the "extern i/o test".

extern input/output test

If this test should not be carried out (see 7.1), quit block 4 pressing twice.

Remove test plug.

diagnose press key 0...9

8 RS 232 test

A 3.496.8560 test plug normally used in the repair service is required for this test. However, this plug can also

be purchased by customers under the above number.

For the sake of completeness, the procedure is described here. If a diagnostic test of the RS 232 interface is

not required, continue with point 9.

2

3

8

20 6

4

5

Fig. 4

Connections in the 3.496.8560 plug


Press <5>

.

RS232 test 1...2