5.1 Setting up and connecting the instruments

784 KFP Titrino

117

5.1.6 Connection of the 774 Oven Sample Processor

The Oven Sample Processor heats the sample and transfers the
moisture from the sample to the titration vessel of the Titino. Titrino
and Oven Sample Processor are connected via the remote sockets
(cable 6.2141.020) as well as via the RS interfaces (cable
6.2125.110)

774

The "Remote" socket allows not only connection of a sample
changer but also additional control functions. Pin assignment of
the "Remote" socket and control possibilities, see page 125.

In connections with the oven sample processor, "auto start"
should be set to "OFF" in the key. The start com-
mand is given by the oven sample processor.

In addition, while a sequence is being processed, the 774 Oven
Sample Processor can, via the serial RS interface, cause the 784 Ti-
trino to load a particular method. The Titrino can obtain the oven
parameters via the RS connection.

C

B

A

D

784

6.2141.020 Cable (Remote)

6.2125.110 Cable (RS 232)