2.4 Configuration, key

784 KFP Titrino

12

>RS232 settings COM1

baud rate: 9600

data bit: 8

stop bit: 1

parity: none

handshake: HWs

Settings of RS232 interface

see also page 90 Identical for COM2.

Baud rate (300, 600, 1200, 2400, 4800, 9600, 19200,

38400, 57600, 115200)

Data bit (7, 8)

Stop bit (1, 2)

Parity (even, odd, none)

Handshake (HWs, SWline, SWchar, none)

see page 90.

>common variables

C30 0.0
etc.

Values of the common variables

Common variables C30...C39 (0..

± 999999)

The values of all common variables are displayed. For

creating common variables see page 30.

Settings with key and power ON
Proceed as follows:
1. Switch the Titrino OFF.
2. Press and keep it pressed during switching the Titrino ON.

The display shows the following:

lock:

Locking keys , and

DATA>, and the functions "recall method", "store

method" and "delete method" of the internal method

memory in the Titrino.

curve:

Changes the appearance of the curve printout.

>lock

: OFF
: OFF
: OFF

recall method: OFF
store method: OFF
delete method: OFF

lock

"ON" means that the corresponding function is not

accessible anymore.

The corresponding key is locked.

The corresponding function in the internal method

memory of the Titrino is locked.