Metrohm 784 KFP Titrino User Manual
Page 122

5.1 Setting up and connecting the instruments
784 KFP Titrino
119
Settings at the Titrino:
Under key
Barcode:
input
The barcode string goes to the entry field in which the cursor is currently lo-
cated.
method
If the silo memory is switched on the barcode string always goes to the
method. The cursor position has no effect.
If the silo memory is switched off the input has no meaning.
id1
The barcode string always goes to Id1. The cursor position has no effect.
id2, id3
As for id1.
smpl size
The barcode string always goes to the sample size. The cursor position has no
effect. If the silo memory is switched on the silo line will be concluded with the
sample size and the cursor moves to the next silo line.
Settings at the barcode reader:
Plug the barcode reader into the Remote Box. The barcode reader instruction manual
contains the codes which you must enter.
1. Bring the barcode reader into the programming mode.
2. Make the necessary setting for emulating a PC keyboard (may be country-specific).
Select
3. Exit the programming mode.
Notes:
•
If longer characters chains than are permitted by the corresponding input are trans-
mitted then the first n characters will be accepted; the last characters will be cut off.
•
If the silo memory is switched on and the settings "barcode: method" or "barcode: idX"
are operative, the first silo line will be created when the string is received. Higher silo
lines than 1 are only created and concluded with the sample size.
5.1.8.2 Connecting a PC keyboard
PC keyboards with a 5-pole DIN plug can be connected to 6.2148.000 Remote Box. For
keyboards with a PS/2 plug an adapter PS/2
→
DIN is available in PC shops.
Settings at the Titrino:
Under key
Keyboard:
Select the country-specific keyboard layout of your PC keyboard.
If the Titrino does not support your keyboard you should select a keyboard which has the
closest possible layout (for example check the 2nd occupancy of the numerical keys).
Country-specific special characters will probably not be converted correctly.