Metrohm 784 KFP Titrino User Manual
Page 21

2.6 Parameters, key
784 KFP Titrino
18
>stop conditions
stop V: abs.
**titr.
stop V 99.99 ml
**titr.
factor 999999
**titr.
filling rate max. ml/min
**titr.
Stop conditions for titration
If this is not "normal", i.e. after reaching the EP
Type of stop volume (abs., rel., OFF)
"abs.": absolute stop volume in mL
"rel.":
relative stop volume to sample size
"OFF": stop volume switched off. Stop volume is not
monitored.
The stop volume applies for one single titration. The
conditioning volume is not monitored.
If "abs." is set:
Absolute stop volume (0...9999.99 mL)
If "rel." is set:
Factor for relative stop volume (0...±999999)
Calculated as: Stop V in mL = factor
∗
sample size
Filling rate after titration (0.01...150 mL/min, max.)
The maximum rate depends on the Exchange Unit:
Exchange Unit
max. rate
5 mL
15 mL/min
10 mL
30 mL/min
20 mL
60 mL/min
50 mL
150 mL/min
>preselections
conditioning: ON
display drift: ON
cond.
drift corr: OFF
cond.
drift value 0.0
µl/min
cond.
req.ident: OFF
cond.
Preselections for the sequence
Automatic conditioning of titration vessel (ON, OFF)
If conditioning is "ON", the solution is constantly kept at
the endpoint.
When conditioning is performed, the volume drift can
be displayed during the conditioning:
Display of drift during conditioning (ON, OFF).
Volume drift.
Type of drift correction (auto, man., OFF)
Type of drift correction: (EP – drift
∗
time)
auto: drift value at start is valid
man.: see below
OFF : no correction
Value for manual drift correction (0.0...99.9
µL/min)
Request of identifications after start of titration (id1,
id1&2, all, OFF)
After start, sample identifications can be requested
automatically: Only id1, id1 & id2, all three id's or no
inquiries.