2.4 Configuration, key
784 KFP Titrino
11
barcode: input
Target for barcode reader (input, method, id1, id2, id3,
smpl size)
The barcode reader is used as an input aid, see
page 118.
Input:
The barcode string goes to the entry field in
which the cursor is currently located.
Method:
The barcode string goes to the entry field
"Methods" in the silo memory.
Id1:
The barcode string goes to the entry field
"Id1". (Similar for Id2 and Id3.)
Smpl size: The barcode string goes to the entry
field "smpl size".
>auxiliaries
dialog: english
date 1999-08-15
time 08:13
run number 0
auto start OFF
start delay 0 s
result display: bold
dev.label.
program 784.0010
General settings
Selection of dialog language (english, deutsch,
français, español, italiano, portugese, svenska)
Current date (YYYY-MM-DD)
Format: Year-month-day, entry with leading zeros.
Current time (hh-mm)
Format: Hours-minutes, entry with leading zeros.
Current run number for result output (0...9999)
The sample number is set to 0 when the instrument is
switched on and incremented on every determination.
Automatic start of titrations. (1...9999, OFF)
Number of automatic starts ("number of samples").
Used for instrument interconnections in which the
external instrument does not initiate a start. Not
advisable in connections with Sample Changers.
Start delay (0...999999 s)
Delay time after start of methods. Abort start delay time
with
Type of result display at the end of the determination
(bold, standard)
bold: The calculated results are displayed in bold
characters.
standard: Displays the whole information, e.g. results,
endpoints, messages etc.
Individual identification of devices (up to 8 ASCII
characters)
Will be printed in the result report, see page 32.
Display of program version