2.6 Parameters, key

784 KFP Titrino

15

2.6 Parameters, key

The key is used for the entry of values that
determine the modes. Values marked with "cond." are
accessible during the conditioning in the KFT mode.
"**titr." means that these values can be changed dur-
ing the titration. They influence the ongoing determina-
tion. Other values can only be changed in the inactive
state.
The display texts of the Titrino are shown on the left
side. The values are the default values.

2.6.1 Parameters for KFT

control parameters:
for the EP.
titration parameters
control the general course of the titration.
stop conditions:
Parameters for the termination of the titration.
statistics:
Calculation of mean values and standard deviation,
see page 28.
preselections:
ON/OFF of various auxiliary functions such as auto-
matic requests after the start and activate pulse.

>control parameters

EP at U 250 mV
cond
.

dynamics 100 mV
**titr.

Control parameters

Endpoint (input range depending on the measured

quantity:

U, Ipol:

0.. ±2000 mV

I, Upol:

0...±200.0

µA)

Control (input range depends on the measured

quantity:

U, Ipol:

1...2000 mV

I, Upol:

0.1...200.0

µA)

Outside of the control range: constant dosing, see

page 21.

parameters

>control parameters

>titration parameters
>stop conditions
>statistics
>preselections

PARAM