Metrohm 784 KFP Titrino User Manual
Page 20

2.6 Parameters, key
784 KFP Titrino
17
start V: OFF
cond.
start V 0.0 ml
cond.
factor 0
cond.
dos.rate max. ml/min
**titr.
pause 2 0 s
**titr.
extr.time 0 s
**titr.
I(pol) 50
µA
U(pol) 400 mV
electrode test: OFF
temperature 25.0 °C
cond.
time interval 2 s
cond.
Type of start volume (OFF, abs., rel.)
"OFF": start volume switched off
"abs.": absolute start volume in mL
"rel.":
relative start volume to sample size.
If "abs." is set:
Absolute start volume (0...999.99 mL)
If "rel." is set:
Factor for relative start volume (0...±999999).
Calculated as: start V in mL = factor
∗
sample size
Dosing rate for start volume (0.01...150 mL/min, max.)
The maximum rate depends on the Exchange Unit:
Exchange Unit
max. rate
5 mL
15 mL/min
10 mL
30 mL/min
20 mL
60 mL/min
50 mL
150 mL/min
Pause 2 (0...999999 s)
Waiting time after start volume, e.g. reaction time after
dosing of a start volume. The waiting time can be
aborted with
Extraction time (0...999999 s)
During this time the titration is running. It will not be
stopped (also if EP is already reached), before the
extraction time is terminated. The extraction time can
be aborted with
Polarization current (–127...127
µA)
or the
polarization potential (–1270...1270 mV, in steps of
10 mV) is inquired.
Electrode test (OFF, ON)
Test for polarized electrodes. Performed on
changeover from the inactive standby state to a
measurement. "OFF" means that the test is not
performed.
Titration temperature (–170.0...500.0 °C).
The temperature can be entered manually.
Time interval (1...999999 s)
Time interval for acquisition of a measured value into
the measuring point list.