beautypg.com

6 appendix, 3 user methods, 1 general 6.3.2 kf titer determination with h – Metrohm 784 KFP Titrino User Manual

Page 126: O or methanol standard "h2otiter, 4 titrino validation, 5 warranty and certificates, 6 scope of delivery and ordering designations, 1 technical specifications

background image

6.1 Technical specifications

784 KFP Titrino

123

6 Appendix

6.1 Technical specifications

Modes

KFT: Karl Fisher Titration for water determination

Measuring inputs

1 measuring input for polarized electrodes.

Measuring range

Voltage

0...±2000 mV, resolution 1 mV, error limit 0.1 % full scale

Current

0...±200.0

µ

A, resolution 1

µ

A

Water content

a few ppm to 100 %

Polarizer

Ipol: 0...±127

µ

A, resolution 1

µ

A

Upol: 0...±1270 mV, in steps of 10 mV

Dosification

Volume of buret cylinder 1, 5, 10, 20 or 50 mL
Resolution

10 000 steps per buret cylinder

Titrating burets

1 internal buret

Auxiliary burets

1 additional buret: 765 or 776 Dosimat

Materials

Housing

Polybutyleneterephthalate (PBTP)

Keypad cover

Polycarbonate (PC)

Display

Graphical LCD, 192 x 64 Dots
Field: 100 x 37 mm
LED back-lit

Memory

Method memory for up to 100 methods
Data bank with 18 Metrohm methods
Silo memory for sample data and results

RS232 interface

2 separate interfaces, each can be configured
for printer, balance or computer connection: completely
controllable from external control unit

Remote input/output lines

for Sample Changer, robot connection, oven, ultra turax...
With optional Remote Box:
Connection of barcode reader and PC keyboard.