2.10 Data output

784 KFP Titrino

32

Original reports which are put out automatically at the end of the titration can be printed

with recalculated values at any time. Key sequence:

.

Original reports have double dashes ==== at the end, whereas recalculations are

marked by single dashes ----.

Report outputs can be stopped with .

Example of reports:

Full result report

User (only if entered)

Device label (if there is a designation, see

page 11) and manual signature.

Curve

Scaling of time axis

'fr
784 KFP Titrino 02134 784.0010
user sn
date 1999-08-15 time 15:01 3
KFT Ipol KF
smpl size 0.879 g
EP1 2.5725 ml
Water 1.44 %
Titer 4.9372 mg/ml
mean( 3) +/–s s/%
Water 1.46 0.027 % 1.86
device label Titr 1 sign:
============

'cu
784 KFP Titrino 02134 784.0010
user sn
date 1999-08-15 time 15:01 3
KFT Ipol KF
5.0 s/div dV=0.5 ml/div

============