5.1 Setting up and connecting the instruments
784 KFP Titrino
116
If you enter one of the COMs of the Titrino for this parameter then
your Titrino result report will contain the oven data "heating time",
"sample temp.", "lowest temp.", "highest temp." and "gas flow". The
start is triggered at the oven. When the Titrtrino titration vessel has
been conditioned the oven automatically starts the titration.
The 707 KF Oven can also be connected instead of the 768 KF
Oven.
5.1.5 Connection of a Sample Changer
With 6.2141.030 cable (instead of 6.2141.020), two Titrinos can be
connected to the 730 (or 760) Sample Changer at the same time.
•
The "Remote" socket allows not only connection of a sample
changer but also additional control functions. Pin assignment of
the "Remote" socket and control possibilities, see page 125.
•
If a calibration has to be performed with the sample changer, the
calibration parameter "sample changer:" must be set to "ON".
•
In connections with the sample changer, "auto start" should be
set to "OFF" in the
by the Sample Changer.
730
C
B
A
D
784
6.2141.020 Cable