beautypg.com

Metrohm 784 KFP Titrino User Manual

Page 129

background image

6.2 Pin assignment of the "Remote" socket

784 KFP Titrino

126

Outputs

pin 17 (Output 3)

pin 3 (Output 4)

pin 16 (Output 5)

pin 1 (Output 6)

pin 2 (Output 7)

pin 6 (Output 8)

pin 7 (Output 9)

pin 8 (Output 10)

pin 13 (Output 11)

pin 19 (Output 12)

pin 20 (Output 13)

End of determination

EOD

not used

Error,

active with errors

Activate pulse, see page 128.

L6 in TIP

Pulses for recorder (t

p

=150

µ

s)

10 000 per buret cylinder

not used

not used

not used

not used

sample size out of limits

result out of limits, see page 26

For all outputs:

V

CE0

= 40 V

I

C

= 20 mA

t

Pulse

> 100 ms

Functions see page 127.

Voltage

pin 15

pin 14

pin 25

I

200 mA

0 V: active

5 V: inactive

Ordering numbers for plug:
K.210.9004 (shell) and K.210.002

No liability whatsoever will be accepted for damage caused by im-
proper interconnection of instruments.

0 V

+5 V