4.1 Error and special messages
784 KFP Titrino
98
error 45
The receive buffer of the Titrino contains an
incomplete string (missing L
F
). Transmission of the
Titrino is thus blocked.
Exit: Send L
F
or
manual stop
The determination has been manually stopped.
meas.pt list overflow
Maximum 500 measured points can be stored.
Exit: Use start criteria or select larger time interval.
missing EP
An EP needed for calculation in a formula is missing.
no new com.var.
The common variable could not be assigned as the result or
the mean value could not be calculated. The old value remains
in force.
no new mean
No new mean value has been calculated as at least one
quantity stipulated for mean value calculations could not be
calculated.
no new silo result
No new silo result C24 or C25 could be stored as the assigned
quantity could not be calculated.
no oven param.
The oven could not be found at the given COM.
Remedy: connect the oven to the given RS-interface of the Ti-
trino or set the following in your method under ,
>preselections, "Oven: no".
no titration data
No curve can be printed as no data are available.
not valid
A value is not available.
overrange
The measuring range of
±
2 V has been exceeded. Overrange
replaces the corresponding measured value (U or I). If a
measured value is in overrange (primary or secondary
measured value), the other (secondary or primary measured
value) can also be unstable.
result out of limits
The result lies outside the limits which were defined in the
method, see page 26.
Exit: Calculate result again or new start.
sample size out
The sample size is outside the limits which are defined in the
method, see page 19.
Exit: Enter new sample size.
service is due
The service interval has elapsed. Contact Metrohm service so
that the Titrino can be serviced. This message will appear
each time the Titrino is switched on.
Exit: New start.
silo empty
The silo memory is switched in but empty and a titration has
been started. Corrective action: At least the first 1 silo line
before starting the first titration.
Exit:
silo full
The silo memory is full up. Corrective action: If you have filled
less than 255 silo lines, you can create more space by
deleting old methods no longer needed. 1 silo line needs
18...120 bytes.