beautypg.com

Metrohm 784 KFP Titrino User Manual

Page 30

background image

2.7 Result calculations

784 KFP Titrino

27

Input of method specific operands C01...C19, key

With the operands C01...C19 can be put

in. For the calculation, the operands are used, which

were introduced in the formula.

The inputs method specific and are store in method
memory.

The calculation report can be printed with the key sequence

<

/

> (press keys repeatedly until "calc" appears in the display) .

Example for a calculation report:

Calculation report

mode und method name

formulas:

result name=formula;number of decimal

places;result unit

sample size

method specific operands

values of the common variables that are

used

C-FMLA

C

1

'fm
784 KFP Titrino 02134 784.0010
user sn
date 1999-08-15 time 15:01 3
KFT Ipol KF-Blank
>calculations
Water=(EP1-C38)*C39*C01/C00/C02;2;%
Titer=C39;4;mg/ml
Blank=C38;4;ml
C00= 1.0
C01= 0.1
C02= 1
C38= 0.0
C39= 0.0
------------