7 various functions, 1 calibration and addition/subtraction data, 1 ph calibration data – Metrohm 781 pH/Ion Meter User Manual
Page 95: Various functions, Calibration and addition/subtraction data, Ph calibration data, 7various functions

7 Various functions
780/781 pH/Ion Meter, Manual
85
7
Various functions
This section describes various functions of the pH/Ion Meter which can-
not be assigned to individual modes.
7.1
Calibration and addition/subtraction data
The current pH calibration data can be called up at any time with the
concentration calibration (
tion data (
A detailed description of the evaluation of the calibration and addi-
tion/subtraction measurements for the calculation of this data can be
found in Section 9.2.
An overview appears which, depending on the measuring mode and
calibration data, can have a different composition.
For example, if no calibration has yet been carried out for a new elec-
trode Id then only the theoretical calibration data slope = 100.00% and
pH(0) = 7.000 will be shown and taken into account for the measure-
ment.
7.1.1
pH calibration data
CAL.DATA
7
CAL.DATA
├
electr.id
├
method
├
slope
├
pH(0)
├
temperature
├
temp.sensor
├
c.dat.
├
cal.interval
├
variance
├
buffer type
├
no.of buffers
├
cal.table
├
b1
├
b2
:
├
bn
└
curve
electr.id
If an electrode Id has been selected or entered under
PARAMETER pH/
measuring parameters/electr.id
before the calibration then it will
be shown here. The calibration data apply only to this electrode and will
change when a new electrode Id is set under the above measuring pa-