Metrohm 781 pH/Ion Meter User Manual
Page 36

3.2 Determination of an ion concentration (only 781: Conc mode)
26
780/781 pH/Ion Meter, Manual
5 Poor calibration data
•
If the calibration data lies outside certain limits
(slope ≤ 80% or ≥ 105% of the theoretical slope of 59.16 mV
(25°C) per ionic charge) or the sign of the slope does not
agree with the selected ion type a corresponding message will
be shown:
•
You can still accept the calibration data with
ject it with
6 Show calibration data
•
You can now display the calibration data with
If you move the selection bar to the end of the list with <
>,
then in the line
Curve < -> >
you can activate the calibration curve display with <->>.
7 Print calibration report
•
With
printer or a PC.
•
With
tion report in the report selection list (calib short); this contains
all the data except the calibration curve itself.
•
The Select key <->> offers the further option '
calib full
';
this also outputs the calibration curve (see Section 7.2.4).
ENTER
MODE
CAL.DATA
7
REPORT