beautypg.com

Metrohm 781 pH/Ion Meter User Manual

Page 36

background image

3.2 Determination of an ion concentration (only 781: Conc mode)

26

780/781 pH/Ion Meter, Manual

5 Poor calibration data

If the calibration data lies outside certain limits

(slope ≤ 80% or ≥ 105% of the theoretical slope of 59.16 mV

(25°C) per ionic charge) or the sign of the slope does not

agree with the selected ion type a corresponding message will

be shown:

You can still accept the calibration data with , or re-

ject it with .

6 Show calibration data

You can now display the calibration data with .

If you move the selection bar to the end of the list with <

>,

then in the line

Curve < -> >

you can activate the calibration curve display with <->>.

7 Print calibration report

With you can now output a calibration report to a

printer or a PC.

With you can jump directly to the short calibra-

tion report in the report selection list (calib short); this contains

all the data except the calibration curve itself.

The Select key <->> offers the further option '

calib full

';

this also outputs the calibration curve (see Section 7.2.4).

ENTER

MODE

CAL.DATA

7

REPORT

9

This manual is related to the following products: