Metrohm 781 pH/Ion Meter User Manual
Page 37

3 Short operating tutorial
780/781 pH/Ion Meter, Manual
27
'fc
781 pH/Ion Meter
01104 5.781.0010
date 2004-05-13 09:20:50
Conc ******** run number
1
electr.Id
conc.calibration
ion type F(-1)
temperature 22.7 °C
temp.sensor Pt1000
c.dat. 2004-05-12 09:32
conc/ppm U/mV dconc/%
std. 1 5.00E-01 139.9 0.2
std. 2 2.50E+00 97.0 0.1
std. 3 5.00E+00 78.8 -0.6
std. 4 2.50E+01 37.1 -0.2
std. 5 5.00E+01 19.2 0.4
variance 0.014
slope -59.9 mV
E(0) 120.8 mV
c(blank) 0.00E+00 ppm
==============