2 preselections, Preselections – Metrohm 781 pH/Ion Meter User Manual
Page 93

6 Methods / Parameters
780/781 pH/Ion Meter, Manual
83
auto dos
increment 1...n 1.0E-30...0.1...999.9 mL
The particular volume increment must be entered for each addi-
tion of the standard or sample solution.
auto
delta U
1...10...999 mV
Under
auto
the required potential difference must be entered.
Please note that the Dosimat Plus must be properly prepared be-
fore the start of an addition run (cylinder freshly filled, air bubbles
removed, etc.).
During an addition you can terminate the run prematurely with it
be used for the calculation of the result. This manual termination will be
marked in the report.
6.6.2
Preselections
PARAMETER Conc
:
└
preselections
├
req.ident
├
req.smpl.size
└
activate pulse
req.ident:
Id1, Id1 & Id2, OFF
For sample identification you can document each measured value with
two Ids in the report header. These can either be predefined under
CONFIG/Report
and included in each report or, depending on the Pre-
selections setting, they can be requested before each measurement.
Please note that sample identifications requested as a preselection be-
fore the measurement will also be stored as an entry under
CON-
FIG/Report
. This means that they are again available for the next
measurement.
req.smpl.size : value, unit, all, OFF
The sample size and sample size unit parameters described under the
calculation parameters (see Section 6.5.4) can also be requested be-
fore each measurement. This allows a more individual input of the
sample data.