8.6 pH electrode test

132

780/781 pH/Ion Meter, Manual

Test criterion

Message

Measure

pH 9: absolute drift in stirred

solution > 1 mV/min

problem

Clean diaphragm and/or

regenerate the glass membrane.

pH 9:

-10 mV < U(unstirred) < 10 mV

AND

sum of abs. drift values after 1, 2,

3 and 4 min < 12 mV/min

short circuit

Replace electrode (short circuit

or crack in glass membrane).

2 slopes do not meet the

conditions for a passing

electrode

check buffer/
electrode

Repeat test using correct

buffers.

Electrode poor as streaming

potential dU is too high

bad diaphragm

Clean diaphragm.

All slopes meet conditions for

passing electrode

AND

Uoff outside the preset limit

reference not
matching

Repeat with suitable reference

system; replace contaminated

reference electrolyte. If

necessary select electrode type

'

own

' and adapt the Uoff limits.

All slopes do not meet conditions

for passing electrode

partial short
circuit

Check temperature sensor or

enter correct temperature. If this

does not help then replace pH

electrode.

A response time does not meet

the criterion for passing electrode

glass
membrane/
reference

Regenerate glass membrane

and/or clean diaphragm.

8.6.5

Care and maintenance of pH glass electrodes

General

A combined pH electrode must be filled with the correct refer-

ence electrolyte, e.g. with c(KCl) = 3 mol/L.

The reference electrolyte chamber should be completely filled

with clean electrolyte.

The reference electrolyte filling opening must remain open during

the measurement and be closed during storage.

There should be no air bubbles in the inner or reference electro-

lyte.

All cable connections should be clean and dry.