8 Troubleshooting – Messages – Maintenance

780/781 pH/Ion Meter, Manual

115

Problem

Possible cause

Remedy

With standard addition
'auto' the potential step
'delta U' is not reached

Streaming potential of the
measuring or reference elec-
trode causes potential differ-
ence between addition in stirred
and measurement in unstirred
solution

Lower dosing rate

Increase stirring rate

Increase 'delta U'

'auto dos' instead of 'auto'

System test message:
system error #X

Instrument fault

Note error number and inform

Metrohm Service

8.2

Messages

The pH/Ion Meter uses various messages to selectively inform you

about possible errors or operating problems. These are shown in the

bottom line of the display and illustrated by a symbol in front:

Type

Symbol

Example of a message

Error

same buffer

Warning

cal.data out of limits

Information

manual stop

change buffer

calibration OK

Monitoring

cal.interval expired


These messages can normally be removed by pressing the

key. However, some messages will remain until their cause has been

remedied. For example, this applies to a monitoring message or a

warning message during a standard addition run. In the latter case it is

necessary to carry out a new and successful analysis to remove such a

warning message.

Please note that you should not press to acknowledge a

message while a report is being produced as this will cancel the re-

port.

The following table lists all the messages which could appear on the

display of the pH/Ion Meter.