2 report identification, 3 measuring point report, Report identification – Metrohm 781 pH/Ion Meter User Manual
Page 109: Measuring point report

7 Various functions
780/781 pH/Ion Meter, Manual
99
7.2.2
Report identification
The report identification, known here as Report Id, is an abbreviation
describing the type of report and is primarily important for the remote
control of the pH/Ion Meter via the RS232 interface. Computer pro-
grams such as Metrohm Vesuv
©
3.0 for Windows
TM
can identify the
produced report by this identification and use it for further evaluation.
The following report Ids exist for the 780 pH Meter and 781 pH/Ion Me-
ter:
Report Id
Report name
Availability
'mp
measuring points
780, 781
'co
configuration
780, 781
'pa
parameter
780, 781
'um
user methods
780, 781
'mv
measured values stored
780, 781
'cr
pH calibration report
780, 781
'fp
full pH calibration
780, 781
'di
diagnose
780, 781
'et
electrode test
780
'cc
conc. calibration
781
'fc
full conc. calibration
781
'ca
conc. addition
781
'fa
full conc. addition
781
7.2.3
Measuring point report
By pressing the
urement types a measuring point report will be produced. The following
examples show versions of this type of report, the various configura-
tions of which are defined under
CONFIG/print meas.value
(see
Section 5.2).
A typical version of a measuring point report is the output of a single
measured value. The print criterion '
immediate
' or '
Drift
' must be se-
lected.
'mp
781 pH/Ion Meter
01104 5.781.0020
date 2004-05-28 17:00:56
user
C. Weber
pH pH-1 run number
1
electr.id
id1 charge 21
id2 sample A01
#1 pH = 8.182
21.2 °C (Pt1000)
2004-05-28 17:00:56
==============
MEAS/