beautypg.com

Metrohm 781 pH/Ion Meter User Manual

Page 35

background image

3 Short operating tutorial

780/781 pH/Ion Meter, Manual

25

2 Start calibration with first standard

Immerse electrodes in the standard with the lowest F

-

concen-

tration

Start calibration with

If no temperature sensor is connected:

enter the temperature and exit with

Confirm standard concentration with or – if neces-

sary– correct it

Wait for measurement of the potential

3 Continue calibration with increasing standard con-

centrations

When the request 'Change std. ' appears, remove

the electrodes from the first standard, rinse them with water

and blot them gently with a clean lint-free tissue

Immerse electrodes in the next standard and continue the cal-

ibration sequence with

In each case conform the standard concentration with

or – if necessary – correct it

Repeat procedure for each remaining standard

4 Result

Wait for potential measurement, result will be displayed:

After 30 s the pH/Ion Meter will switch back automatically to

the measurement display. This can also be done immediately

with .

CAL

ENTER

ENTER

This manual is related to the following products: