9.2 Evaluation
140
780/781 pH/Ion Meter, Manual
The calculation of the calibration data depends on the number of pH
buffers used:
1-Point calibration
Slope:
The theoretical slope value of 100.0% is used.
pH(0)
=
N
1
1
U
Slope
U
pH
⋅
−
−
2-Point calibration
Slope
=
(
)
(
)
N
1
2
1
2
U
1
pH
pH
U
U
−
⋅
−
−
pH(0)
=
N
U
Slope
U
pH
⋅
−
−
;pH , U = mean value
3- and more point calibration
With at least 3 calibration buffers a compensation curve according to
the principles of the smallest mean-square error (linear regression) will
be calculated.
Slope
=
(
) (
)
(
)
N
i
2
i
i
i
i
U
1
pH
pH
U
U
pH
pH
−
⋅
−
−
⋅
−
∑
∑
pH(0)
=
N
U
Slope
U
pH
⋅
−
−
Variance =
(
)
2
n
U
U
i
2
calculated
,i
i
−
−
∑
;n = no. of measuring points