Metrohm 781 pH/Ion Meter User Manual

Page 128

background image

8.2 Messages

118

780/781 pH/Ion Meter, Manual

Message

Situation

Cause

Remedy

buffers
unsuitable

Electrode test

Electrode test has been
started with a set of buffers
that does not contain all
the necessary buffers

Change buffers

Check selected buffer

type in the calibration pa-
rameters (see Sec-
tion
8.6.1)

change buffer

pH calibration

Instrument is waiting for
new buffer

1. Change buffer
2. Press

buffer alloc.
impossible?

pH calibration

Buffer not recognized or

not defined (buffer type
'special', 'own' or 'mixed')




Offset potential Uoff is

set too high

1. Replace buffer
2. Press

or
1. Stop calibration
2. Define buffer
3. Restart calibration
or
Enter Uoff (see

Sec-

tion 6.2.2

) and restart cali-

bration

report output
...

Standard addi-
tion,
sample addition

Instrument is busy with the
output of a RS232 report

Wait until report output is
finished

RS error 36

RS232 remote
control

RS receiver error:
parity unmatched

1. Press

2. Match the parities of both

instruments (see

Sec-

tion 5.7

)

RS error 37

RS232 remote
control

RS receiver error:
stop bit unmatched

1. Press

2. Match the stop bits of

both instruments (see

Section 5.7

)

RS error 38

RS232 remote
control

RS receiver error:
data overflow, at least one
character cannot be read
(incorrect baud rate)

1. Press

2. Match the baud rates of

both instruments (see

Section 5.7

)

RS error 39

RS232 remote
control

RS receiver error:
internal receiver buffer
overflow

1. Press

2. Repeat process

RS error 42

Report output

RS transmitter error:
handshake not replied to
for at least one second

1. Press

2. Check reception readi-

ness at other end

RS error 43

Report output

RS transmitter error:
transmission from the pH
Meter has been interrupted
by XOFF for at least three
seconds

Press

Send XON