Metrohm 781 pH/Ion Meter User Manual
Page 128

8.2 Messages
118
780/781 pH/Ion Meter, Manual
Message
Situation
Cause
Remedy
buffers
unsuitable
Electrode test
Electrode test has been
started with a set of buffers
that does not contain all
the necessary buffers
•
Change buffers
•
Check selected buffer
type in the calibration pa-
rameters (see Sec-
tion 8.6.1)
change buffer
pH calibration
Instrument is waiting for
new buffer
1. Change buffer
2. Press
buffer alloc.
impossible?
pH calibration
•
Buffer not recognized or
not defined (buffer type
'special', 'own' or 'mixed')
•
Offset potential Uoff is
set too high
1. Replace buffer
2. Press
or
1. Stop calibration
2. Define buffer
3. Restart calibration
or
Enter Uoff (see
Sec-
) and restart cali-
bration
report output
...
Standard addi-
tion,
sample addition
Instrument is busy with the
output of a RS232 report
Wait until report output is
finished
RS error 36
RS232 remote
control
RS receiver error:
parity unmatched
1. Press
2. Match the parities of both
instruments (see
Sec-
RS error 37
RS232 remote
control
RS receiver error:
stop bit unmatched
1. Press
2. Match the stop bits of
both instruments (see
RS error 38
RS232 remote
control
RS receiver error:
data overflow, at least one
character cannot be read
(incorrect baud rate)
1. Press
2. Match the baud rates of
both instruments (see
RS error 39
RS232 remote
control
RS receiver error:
internal receiver buffer
overflow
1. Press
2. Repeat process
RS error 42
Report output
RS transmitter error:
handshake not replied to
for at least one second
1. Press
2. Check reception readi-
ness at other end
RS error 43
Report output
RS transmitter error:
transmission from the pH
Meter has been interrupted
by XOFF for at least three
seconds
•
Press
•
Send XON