4 meas tc cond - additional measured values – Metrohm tiamo 2.1 Manual User Manual
Page 968
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
954
■■■■■■■■
tiamo 2.1
Range
0 ... 999999 s
Default value
3000 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
712
Range
0.4 ... 999999 s
Default value
2.0 s
856
Range
0.1 ... 999999 s
Default value
2.0 s
Sample solution
Selection of the sample solution from the table Sample solutions (TC
conductivity).
Selection
DIN | Table entries
Default value
DIN
Start temperatur
Temperature at the time of the start of the measurement.
Range
-20.0 ... 150.0 °C
Default value
20.0 °C
Stop temperature
Temperature at the time of the end of the measurement.
856
Range
-20.0 ... 150.0 °C
Default value
40.0 °C
5.6.4.10.4
MEAS TC Cond - Additional measured values
Tab: Method
▶ MEAS U ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Input
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together with the measured values present in the
default settings in additional measured value columns Calculated # or
External # in the measuring point list .