5 Method
tiamo 2.1
■■■■■■■■
597
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Titrino
Range
0.00 ... 999.99 mL
Default value
0.00 mL
Dosing rate
Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
Start measured value
Note
Is displayed only for Titrando and 855.
Start measured value
Once the start measured value is reached, the dosing of the start volume is
stopped and the next start condition is processed or the titration is started.
If the start measured value is achieved by the addition of a start volume,
then titration will start directly.
Range
-2000.0 ... 2000.0 mV
Selection
off
Default value
off
Dosing rate
Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.
Range
0.01 ... 166.00 mL/min
Default value
5.00 mL/min