Metrohm tiamo 2.1 Manual User Manual
Page 53
■■■■■■■■■■■■■■■■■■■■■■
2 General program functions
tiamo 2.1
■■■■■■■■
39
Identifica-
tion
Description
Commands
.RE{x}.DRC
Correlation coefficient for mean
dosing rate in window x (1…9)
STAT
.RE{x}.DRS
Standard deviation for mean dosing
rate in window x (1…9) in mL/min
STAT
.RE{x}.DRT
Mean dosing rate in window x (1…
9) in mL/min
STAT
.RE{x}.RWL
Lower limit of the evaluation win-
dow x (1…9) in s
STAT
.RE{x}.RWH
Upper limit of the evaluation win-
dow x (1…9) in s
STAT
.RES
Calculated result of the standard
addition in the selected unit
STDADD
.RPO
Current rack position (entry at time
of ending the command); 0 means
'not defined'
MOVE
.RTE
Reference temperature in °C
MEAS Cond
.SAN
Current, absolute swing angle of
the robotic arm in ° (entry upon
ending the command)
MOVE, SWING
.SLO
Electrode slope of the sensor used
for the command (in % for pH sen-
sor or mV for ISE sensor) or elec-
trode slope calculated from the
calibration (for STDADD and CAL
LOOP)
DET pH, DET U,
MET pH, MET U,
SET pH, SET U,
STAT, MEAS pH,
MEAS U, MEAS
T, MEAS Conc,
STDADD, CAL
LOOP, DOS
.SME
Start measured value (measured
value after processing the start con-
ditions) in the unit of the measured
value
DET, MET, SET,
KFT, KFC, BRC,
STAT
.SPO
Current external position (entry
when ending the command); 0
means invalid position
SWING
.STE
Start temperature (temperature
after processing of the start condi-
tions) in °C
DET, MET, SET,
KFT, KFC, BRC,
STAT